Table of Contents

UV Broadband Meters

Broadband meters are among the most used UV meters, as they can be purchased at an affordable price (starting from about 150€). They stand out due to the high reproducibility of the measured values, a high signal to noise ratio, and very easy handling.

All the advantages of broadband meters over spectrometers make them popular for reptile keepers. The interpretation of the measured values is, however, comparably complicated: Broadband meters measure the irradiance weighted with their own sensitivity spectrum, which do not agree with the correspondent name of the broadband meter. Measured values of different lamps or different brodband meters are fundamentally not comparable. Especially a lamp with a lower measured value at one braodband meter can nevertheless have higher physical uvb irradiance or a higher ability to facilitate vitamin D production.

Without knowledge of

the informative value of a broadband meter is quite limited.

Construction and Function of a Broadbandmeter

Notes regarding nomenclature1)

Photodiode

The basis of a braodbandmeter (radiometer) is a semiconductor photodiode. The current through this photodiode changes when a photon (light particle) is absorbed by the semiconductor. But a photodiode does not react to every photon equally! The energy or wavelength of the light has to match the “band gap” of the semiconductor. For visible light usually silicon is used, for the uv range SiC, GaN, diamond or GaAsP are needed. The response of the photodiode can be further constricted by a filter that stops radiation that one does not want to measure.

In photoconductive mode the electric current dependes linearly on the number of photons. Therefore, Photodiodes are a good device to measure the irradiance of a light source.

In the end the properties of a broadbandmeter is contained in its spectral sensitivity curve .

Spectral sensitivity of a UVB Broadband Radiometer: The meter responds with maximum sensitivity to radiation between 280nm and 290nm. Then the sensitivity drops. At 310nm the sensitivity is only 50%.

This sensitivity curve, together with the spectrum of the light source, determines the signal of the broadbandmeter: A lamp with 10µW/cm² in the range between 280nm and 290nm and nowhere else with give the same signal as a lamp with 20µW/cm² in the range of 305nm to 315nm, because the sensitivity of the meter is only 50% at 310nm. On the other hand, the same intensity of uv light (10µW/cm²) at 285nm and 310nm will not give the same signal. The signal at 310nm is only 50% of the signal at 285nm.

This behaviour is expressed in the mathematical formula. The current signal of a photodiode is This formula can be written as a sum () or an integral (), and is probably easier to understand for people with little mathematical training as a sum. The instruction of the formula is: At every wavelength ( can be 280nm, 281nm, 282nm … in steps of 1nm) measure the spectral irradiance of the light source and multiply it with the spectral sensitivity of the meter at this specific wavelength . When done, add the results for all wavelengths.

Example:

1 280 nm 0.002 µW/cm²/nm 100 0.002 µW/cm²/nm
2 285 nm 0.001 µW/cm²/nm 99 0.001 µW/cm²/nm
3 290 nm 0.035 µW/cm²/nm 97 0.034 µW/cm²/nm
4 295 nm 0.049 µW/cm²/nm 93 0.137 µW/cm²/nm
5 300 nm 0.529 µW/cm²/nm 81 0.429 µW/cm²/nm
6 305 nm 1.829 µW/cm²/nm 73 1.638 µW/cm²/nm
7 310 nm 5.477 µW/cm²/nm 65 3.541 µW/cm²/nm
8 315 nm 13.442 µW/cm²/nm 58 7.398 µW/cm²/nm
9 320 nm 13.627 µW/cm²/nm 44 5.998 µW/cm²/nm
10 325 nm 17.569 µW/cm²/nm 21 3.942 µW/cm²/nm
11 330 nm 25.637 µW/cm²/nm 4 1.133 µW/cm²/nm
12 335 nm 32.354 µW/cm²/nm 0 0.043 µW/cm²/nm
13 340 nm 41.639 µW/cm²/nm 0 0.071 µW/cm²/nm
24.36 µW/cm²/nm
5 nm 24.36 µW/cm²/nm = 121.8 µW/cm²
Working principle of a UVB braodband radiometer.

Cosine Correction

A ground glass in front of the detector increases the angular range that is seen by the meter (cosine correction). Without this ground glass, the result would depend too much on the angle of the meter with respect to the light source.

Calibration

The current from the photodiode is then multiplied in the electronics of the meter with a calibration factor to get a usable value for the result. Many broadband meters give the irradiance in µW/cm² or W/m². Others use a weighted irradiance like UV index or vitamin D (international units per minute).

To get this calibration factor, a lamp calibration lamp is used. The lamps is measured with a spectrometer and with the broadbandmeter and the two results are compared. The calibration factor is chosen so, that the two results become equal.

Example

Measurement result of the braodbandmeter
Measurement result of the spectrometer

UVB result spectrometer 140 µW/cm²
current signal UVB radiometer 122 µW/cm²
Calibration faktor 140/121.8 = 1.147

In the end, the measurement value of the radiometer for a light source is:

Calibration methods

There are in general three ways of calibration: Using a line source as calibration lamp, using a broad band light source as calibration lamp or measuring the spectral sensitivity of the meter [120]

A mercury vapour lamp is generally used (lines at 254 nm, 313 nm, 365 nm) as a line source. To determine the correct value the lamp can be measured with a simple power meter. This method is quick, simple and reproducible. When a broadband meter is calibrated using a line source, measuring a broadband light source often gives a too small value, measuring a line source a too high value [119].

For application in the field of medicine, calibration with a broad band light source is used. Here the broad band calibration light source must be characterized with a spectrometer, to find the correct value.

The most substantiated but laborious calibration method is determining the spectral response of the broadband meter.

Correction factor / Systematic error

The calibration factor is chosen so, that the broadband radiometer shows the same result as a spectrometer for the calibration lamp. As soon as the tested lamp has a spectrum that differs from the spectrum of the calibration lamp, the radiometer value will vary from the spectrometer value. This is, because in the above sum, the spectrum of the lamp has a huge influence.

This error is called systematic error: If the spectrum of the lamp is known, this error is not a problem, because one can calculate what value the broadbandmeter should give and what spectrometer result one would expect and calculate a correction factor [38; 119]. Afterwards, one simply multiplies the value of the braodbandmeter with this correction factor and the results are as perfect as if a spectrometer had been used all the time, but without all the problems of a spectrometer (long integration time, not very movable, needs a lot of space).

One can even go one step further and calculate a correction factor for this lamp and this radiometer but for a different action spectrum. Once this correction factor is calculated, one can, in theory, use a UVB radiometer to measure the uv-index (of this specific lamp).

And if it turns out, that the correction factor is equal to 1, one could have saved all the afford of calculating it: If you multiply any number with 1, it does not change.

The correction factor is in two cases close to 1, which means, that one does not really need to calculate it and can use the broadbandmeter directly:

The correction factor has to be calculated for every broadbandmeter ( with calibration factor ) and every application (action spectrum and every lamp ). Stating correction factors without taking the specifications of the braodbandmeter, the application and the light source into account are unsound.

If one wants to reduce the depencende on the lamp, the “integral characterization factor” can be used [112] 2) :

Literature on correction factors

Literatur zu Korrekturfaktoren

Quelle Lichtquelle Messbereich Abweichung
Breitband/Spektrometer
spektraler
Korrekturfaktor
[121] artificial sun UV-Index 1/1.85 - 2.22 several broadband meters
[121] artificial sun dUVA 1/1.3 - 5.51 several broadband meters
[320] VitaminD
[474] VitaminD
[38] uv lamps UV-Index 1/2-5 - 1.6
[38] sun UV-Index 1.2 - 2.2
[277,667] Xenon+filter UVA 1.18 - 1.42
[277,667] Xenon+filter UVB 1/25 - 16
[676] sun (132 spectra) UV-Index R²=99.8% without correction
[676] sun (132 spectra) UV-Index R²= 99.98% after correcting vor SZA and ozon
[119] UV lamps, sun UV-Index 0.7-6.3

Accuracy / Statistical Error

The accuracy of every measurement is limited by several statistical and systematic errors [112]. For broadband meters these are:

Ratio of two broadbandmeter readings

The value from one single broadband meter contains hardly any information about the spectrum of the light-source. By adjusting the distance to the lamp, one can measure the same value from every single uv lamp.

But by comparing the readings from two broadband meters with different spectral sensitivity the knowledge about the lamps spectrum can be expanded. The ratio of the two measured values depends on the form (but not the intensity) of the lamps spectrum and the sensitivity of the two broadband meters

If the two normalized sensitivity spectra of the meters have a difference , the ratio can be written as:

This value can be determined from measurements at various distances from the lamp, if the spectrum does not depend on the distance and the cosine correction of both meters are sufficiently equal. For the statistical evaluation regression analysis offers various methods.

Ratio of Solarmeter 6.5 and Solarmeter 6.2 Readings

Recently, the ratio of Solarmeter 6.5 and Solarmeter 6.2 have become a very popular method for lamp testing.

The spectral sensitivity of the Solarmeter 6.5 () and Solarmeter 6.2 () as well as the difference between the two sensitivites is shown:

The ratio of the values depends strongly on the overlap of the lamps spectrum () with . For a lamp with a lot of radiaton in the range 310-320nm, which is not very effective for Vitamin D synthesis, the ratio is large. If this range contains less radiation and the radiation is more concentrated in the range <310nm, the ratio will be small.

The ratio of Solarmeter 6.2 and Solarmeter 6.5 readings contains information whether the UV radiation is concentrated more in the long-wavelength or short-wavelength range.

This method is, of course, not as exact as a spectrometer measurement, but it can give some estimation of the risks of uv damage to the skin

> 60 <0.016 long wavelength concentration (potentially not vitamin d effective)
50-60 0.016 - 0.02 natural sunlight
< 40 > 0.025 unnatural short wavelength concentration
< 20 > 0.05 very short wavelength concentration
< 13 > 0.078 extreme short wavelength concentration

Broadbandmeters

The company Solartech offers several braodbandmeters. The manufacturer provides the spectral sensitivity of the meters and information about the calibration procedure.

Solarmeter 6.2 (UVB)

Solarmeter 6.5 (UV-Index)

Because of the high overlap between vitamin D spectrum (according to DIN/CIE) and because the meter responds to radiation above 315nm only <4%, and above 320nm only <1% this meter is applicable to estimate whether a lamp will allow for vitamin d synthesis.

Literature

[38] ANTON GUGG-HELMINGER, WOLFGANG DÄHN, STEPHAN FENK & B. ANGELO. Broadband and Spectral Measurement Instrumentation for Photobiological Hazard Evaluation. CIE Expert Symposium on Light and Health. (1st Oct 2004)
[112] Characterizing the Performance of Integral Measuring UV-Meters. 2000. UV News 6. A-1–AA-36
[117] NEIL HARRISON. 2000. Improving the Accuracy of Ultraviolet Radiation Measurement. UV News 5. 18–23
[119] ANTON GUGG-HELMINGER, WOLFGANG DÄHN & STEPHAN FENK. 2002. Manufacturer’s view on UV meters with different action spectra. UV News 7. 20–28
[120] PETRI KÄRHÄ. 2002. Calibration and intercomparison issues with broadband UV meters. UV News 7. 29–34
[121] FRANÇOIS J. CHRISTIAENS & ALAIN CHARDON. 2006. Calibration of UV radiometers is needed to guarantee the relevance of measurements. UV News 8. 14–16
[172] FRANCES M. BAINES. Sep 2007. Does the Ambient Temperature Affect the Readings on a Solarmeter 6.2?
[277,667] Discrepancies Associated with Using UVA and UVB Meters to Measure Output of Sunlight, Solar Simulators, etc. Nov 1991. Orlando: Optronic Labaratories
[280] ROBERT M. SAYRE & LORRAINE H. KLIGMAN. 1992. Discrepancies in the measurement of spectral sources. Photochemistry and Photobiology 55. 141–143
[320] THOMAS C. LARASON. 2001. Avoiding Errors in UV Radiation Measurements. Photonics Spectra
[474] WILLIAM H. GEHRMANN, J.D. HORNER, GARY W. FERGUSON, TAI C. CHEN & MICHAEL F. HOLICK. 2004. A comparison of responses by three broadband radiometers to different ultraviolet-B sources. Zoo Biology 23.4. 355–363
[676] JUKKA LINDGREN, WILLIAM H. GEHRMANN, GARY W. FERGUSON & JOHN E. PINDER. 2008. Measuring Effective Vitamin D3-Producing Ultraviolet B Radiation Using Solartech’s Solarmeter® 6.4 Handheld, UVB Radiometer. Bulletin of the Chicago Herpetological Society 43.4. 57–62

1) Notes regarding nomenclature:
  • Action spectrum of the respective effect (UVB, UVA, UV-Index, Vitamin D …): , without a unit, unsually normalized to )
  • Spectral sensitivity of the meter: , without a unit, unsually normalized to )
  • Spectral irradiance of the light source: , unit µW/cm²/nm
  • Spectral irradiance of the calibration lamp: , unit µW/cm²/nm
  • Correction factor:
  • Calibration factor:
2) In oder to eliminate the influence of the calibration lamp one can assume . In this case a factor can be calculated. and are normalized to area=1. is between 0 (no overlap) and 1 (identical spectra)